[Home ]    
Iranian Journal of Radiation Research
:: Main :: About :: Current Issue :: Archive :: Search :: Submit :: Contact ::
:: Volume 3, Number 3 (12-2005) ::
Back to browse issues page Volume 3, Number 3 ( 12-2005), Pages 129 - 133
XML X-ray spectra calculation for different target-filter of mammograms using MCNP Code Print

Author(s): A.A. Mowlavi *
Physics Department, School of Sciences, Tarbiat Moallem University of Sabzevar, P.O. Box 397, Sabzevar, Iran. amowlavi@sttu.ac.ir
Study Type: Original Research | Subject: Radiation Biology
Article abstract:

 ABSTRACT

Background: An electron beam generated X-ray spectrum consists of characteristic X-ray and continuous bermsstrahlung. The aim of this research is calculating and comparing X-ray spectra for different target filter of mammograms.

 Materials and Methods: Monte Carlo is a very powerful tool to simulate a series of different target-filter assembly in order to calculate the X-ray spectra. MCNP version 4C has been used for simulation set up to calculate X-ray spectra for different target-filter combination of mammograms.

 Results: The spectra of different and the most commonly used target– filter in mammography equipments (Mo-Mo, Mo-Rh, Rh-Rh, W-Rh, Mo-Al and Rh-Al) have been calculated.

 Conclusion: The computational results can be used to select a suitable X-ray spectrum for mammography as well as for computing the absorbed dose. Also, the Monte Carlo result of Mo-Mo is in good agreement with the published results.

KeywordsMonte Carlo simulation, mammography, target- filter combination, X-ray spectra, MCNP code,
Full text [PDF 344 kb]

Send email to the article author

 
 
Back to browse issues page Volume 3, Number 3 ( 12-2005), Pages 129 - 133
Iranian Journal of Radiation Research
Static site map - Persian site map - English site map - Created in : 0.08459 seconds by AWT YEKTAWEB 1.8.3.0 by 498 querry